There were some problems in the vascular pros-theses, which effected the health of the patients. This paper aims to presenting the radial compliance of three kinds of vascular prostheses and the pig's carotid based on the dynamic-simulated condition. The radial compliance of knitted structure vascular prostheses with crimps ranged from 2.52 to 0.94%/100mmHg. It was superior to e-PTFE and woven pros-theses', which ranged from 0.73 to 0.31%/100mmHg and from 0.81 to 0.22%/100mmHg. But comparing with the radial compliance of pig's carotid, which ranged from 7.21 to 10.04%/100mmHg, there was a big gap between them. And also the trend of the radial compliance of vascular prostheses was different with the real pig's carotid. There is a lot of work left to improve the radial compliance. Not only to change the vascular prostheses' compliance, but also to meet with the real vascular.
机译:血管假体存在一些问题,影响了患者的健康。本文旨在根据动态模拟情况,介绍三种血管假体和猪颈动脉的径向顺应性。卷曲的针织结构人工血管的径向顺应性范围为2.52至0.94%/ 100mmHg。它优于e-PTFE和机织假体,其范围从0.73至0.31%/ 100mmHg和0.81至0.22%/ 100mmHg。但是与猪颈动脉的径向顺应性相比(7.21%至10.04%/ 100mmHg),两者之间存在较大差距。而且,血管假体的径向顺应性趋势与真猪的颈动脉不同。为了改善径向顺应性,还有很多工作要做。不仅要改变血管假体的顺应性,而且要与真实的血管相遇。