首页> 外文会议>World Congress on Medical Physics and Biomedical Engineering >Evaluation of Downtime of Linear Accelerators Installed at Radiotherapy Departments in the Czech Republic
【24h】

Evaluation of Downtime of Linear Accelerators Installed at Radiotherapy Departments in the Czech Republic

机译:捷克共和国放射疗法部门安装线性加速器停机时间的评价

获取原文

摘要

The National Radiation Protection Institute (NRPI) performed the study evaluating linear accelerator (linac) unscheduled downtime and other parameters related to linac failure (e.g. treatment cancellations, patients transferred to other linac, patients treated with modified dose fractionation) at radiotherapy departments in the Czech Republic. Thirteen radiotherapy departments with at least one linac (out of 21 departments in the Czech Republic) voluntarily participated in the study covering 29 out of 47 linacs. Downtime was evaluated for a one year period from July of 2016 to June of 2017. The methodology was as follows: NRPI designed the data entry form which was sent electronically to medical physicists at participating radiotherapy departments. Data related to linac failures were filled in. The completed forms were evaluated by NRPI. Unscheduled downtime was defined as time when linac cannot be operated during operating hours. Unscheduled downtime per linac per year ranged from 4 to 222 h (mean = 73 h, median = 61 h). Downtime percentage calculated as a ratio of downtime and total sum of operating hours per year ranged from 0.2 to 7.6% (mean = 2.8%, median = 2.2%). The number of treatment cancellations per linac per year ranged from 0 to 661. Unified methodology enabled objective comparison of linac downtime at particular radiotherapy departments in the Czech Republic. The study confirmed usefulness of having minimally two matched linacs at a department. The results of this study could help radiotherapy departments negotiate better service contract (e.g. agreement on maximum guaranteed downtime).
机译:国家辐射保护研究所(NRPI)进行了研究评估了线性加速器(LINAC)未预定的停机时间和与LinaC失败相关的其他参数(例如治疗取消,患者转移到其他Linac的患者,在捷克放射治疗部门治疗。共和国。十三个放射疗法部门至少有一个Linac(捷克共和国21个部门)自愿参加47个Linacs中的29项学习。 2016年7月至2017年6月评估了停机时间。该方法如下:NRPI设计了数据输入表,该表格在参与放射治疗部门以电子学士电子商物发送到医学物理学。与LinaC故障有关的数据填写。通过NRPI评估完成的形式。未安排的停机时间被定义为LinaC在运行时间内无法运行的时间。每年每年的非划分的停机时间为4至222小时(平均值= 73小时,中位数= 61小时)。停机百分比计算为停机比率和每年工作时间总和的比率范围为0.2%至7.6%(平均值= 2.8%,中位数= 2.2%)。每年每年的治疗数量从0到661。统一的方法能够在捷克共和国的特定放射治疗部门的Linac停机客观比较。该研究证实了在部门在部门进行两种匹配的LINACS的有用性。该研究的结果可以帮助放射治疗部门谈判更好的服务合同(例如,关于最大保证停机时间的协议)。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号