首页> 外文会议>IEEE International Instrumentation and Measurement Technology Conference >Influence of the type and position of the sensor on the precision of impedance glucose measurements
【24h】

Influence of the type and position of the sensor on the precision of impedance glucose measurements

机译:传感器类型和位置对阻抗葡萄糖测量精度的影响

获取原文

摘要

Electrochemistry Impedance Spectroscopy has been proposed as a promising technique for non-invasive glucose monitoring. Although it is possible to find related literature to in-vitro and in-vivo experiments, no clear information is available regarding the precision of the measurements. In this paper, we address the problem of the precision due to different samples of the same sensor type and different sensor positions. We will show that the precision of the measurements is high for different sensor positions (over 96%) but can drop significantly when different sensors are used (around 90%), even when these sensors belong to the same brand.
机译:电化学阻抗谱已被提出作为无创葡萄糖监测的有前途的技术。尽管可以找到与体外和体内实验相关的文献,但尚无关于测量精度的明确信息。在本文中,我们解决了由于相同传感器类型和不同传感器位置的不同样本而导致的精度问题。我们将显示,不同传感器位置的测量精度很高(超过96%),但使用不同传感器(大约90%)时,即使这些传感器属于同一品牌,也会显着下降。