首页> 外文会议>IEEE MTT-S International Microwave Workshop Series on RF and Wireless Technologies for Biomedical and Healthcare Applications >Numerical assessment method for implantable cardiac pacemaker EMI triggered by 10MHz-band wireless power transfer coils
【24h】

Numerical assessment method for implantable cardiac pacemaker EMI triggered by 10MHz-band wireless power transfer coils

机译:10MHz带无线电力传输线圈触发植入心脏起搏器EMI的数值评估方法

获取原文

摘要

A numerical assessment methodology to understand the EMI imposed by magnetic resonance type wireless power transfer systems on active implantable medical devices (implantable cardiac pacemaker/cardioverter defibrillator) is introduced. A numerical estimation model that consists of magnetic resonant coils and a human torso phantom with a pacemaker model is constructed. Numerical simulation based on the finite element method yields the interference voltage induced at the connector of the pacemaker inside the torso phantom. Our example assumes magnetic resonance coils operating in the frequency band of 10 MHz.
机译:介绍了一种数字评估方法,以了解由磁共振型无线电力传输系统上的磁共振型无线电力传输系统(可植入的心脏起搏器/心脏除颤器)施加的EMI。构建了一种由磁共振线圈和具有起搏器模型组成的数值估计模型。基于有限元方法的数值模拟产生躯干幻像中的起搏器连接器的干扰电压。我们的示例假设在10 MHz的频带中操作的磁共振线圈。