摘要

A numerical assessment methodology to understand the EMI imposed by magnetic resonance type wireless power transfer systems on active implantable medical devices (implantable cardiac pacemaker/cardioverter defibrillator) is introduced. A numerical estimation model that consists of magnetic resonant coils and a human torso phantom with a pacemaker model is constructed. Numerical simulation based on the finite element method yields the interference voltage induced at the connector of the pacemaker inside the torso phantom. Our example assumes magnetic resonance coils operating in the frequency band of 10 MHz.
机译:介绍了一种数值评估方法,以了解磁共振类型的无线电力传输系统对有源可植入医疗设备(可植入心脏起搏器/心脏复律除颤器)施加的EMI。构造了一个由磁共振线圈和人体模型组成的数值估计模型,并带有起搏器模型。基于有限元方法的数值模拟会在躯干体模内部的起搏器的连接器处产生干扰电压。我们的示例假设磁共振线圈工作在10 MHz的频带中。