首页> 外文会议>IEEE International Conference on Imaging Systems and Techniques >Evaluation of the accuracy of three-dimensional reconstruction of edentulous model jaw relation based on dental articulator positioning
【24h】

Evaluation of the accuracy of three-dimensional reconstruction of edentulous model jaw relation based on dental articulator positioning

机译:基于咬合架定位的无牙颌模型三维重建的准确性评估

获取原文

摘要

A dental articulator and three-dimensional (3D) scanner were utilized to reconstruct the spatial relationship between upper and lower edentulous jaw gypsum models made from five patients. The 7-axis Faro Edge contact measurement system was used to evaluate the vertical and horizontal distances between jaws. Paired t-test results of vertical distance (0.089±0.428 mm) and horizontal distance (−0.082±0.221 mm) between points on the jaws showed no significant differences (P>0.05). One-Way ANOVA of the same measurement data for the five models also showed no significant differences (P>0.05). With an error of less than 0.1 mm, these measurements met clinical requirements. In conclusion, 3D reconstruction of edentulous jaw gypsum models with accurate jaw relation was made possible by scanning the models in centric relation on a dental articulator.
机译:牙咬合架和三维(3D)扫描仪用于重建由五名患者制作的上下无牙颌石膏模型之间的空间关系。 7轴Faro Edge接触测量系统用于评估钳口之间的垂直和水平距离。颚上各点之间的垂直距离(0.089±0.428 mm)和水平距离(-0.082±0.221 mm)的配对t检验结果无显着差异(P> 0.05)。对于五个模型,相同测量数据的单向方差分析也显示无显着差异(P> 0.05)。这些测量值的误差小于0.1毫米,符合临床要求。总之,通过在咬合架上以中心关系扫描模型,可以实现具有精确下颌关系的无牙颌石膏模型的3D重建。