首页> 外文会议>IEEE MTT-S International Microwave Symposium >SAR and Temperature Elevations in Pacemaker Holders Exposed to EM Fields Produced by MRI Apparatus
【24h】

SAR and Temperature Elevations in Pacemaker Holders Exposed to EM Fields Produced by MRI Apparatus

机译:SAR和温度高度在起搏器架暴露于MRI设备生产的EM场

获取原文

摘要

Specific absorption rate (SAR) and temperature distributions, produced inside a thorax model by a magnetic resonance imaging (MRI) apparatus equipped with a birdcage antenna operating at 64 MHz with 100 W of radiated power, have been studied both experimentally and numerically. A pacemaker equipped with a 60-cm long catheter has then been inserted inside the thorax. The presence of this implant gives rise to peak SAR averaged over 1 mm{sup}3 of about 1800 W/kg close to the catheter tip, while the average SAR is about 1 W/kg, that is the same value obtained in the absence of the pacemaker. This peak SAR gives rise to temperature increments up to 6°Cfor 6 minute MRI investigations. However, lower temperature increments are expected if blood perfusion and thermoregulation are considered.
机译:通过磁共振成像(MRI)装置在胸腔模型中产生的特定吸收率(SAR)和温度分布在具有在具有100W辐射功率的64 MHz的鸟笼天线的鸟笼天线中进行,已经通过实验和数值研究。然后将配备60cm长导管的起搏器,然后插入胸部内。该植入物的存在产生超过1mm {sup} 3的峰SAR,其约1800w / kg靠近导管尖端,而平均sar是约1w / kg,即在缺失中获得的值是相同的值起搏器。该峰SAR产生高达6°C的温度增量6°C的MRI调查。然而,如果考虑血液灌注和热调节,则预期较低的温度增量。