首页> 外文会议>IEEE International Symposium on Medical Measurements and Applications >Experimental investigations of viscous heating effect of thermocouples under focused ultrasound applications
【24h】

Experimental investigations of viscous heating effect of thermocouples under focused ultrasound applications

机译:聚焦超声应用中热电偶粘性加热效应的实验研究

获取原文

摘要

Realization of temperature characterization on focused ultrasound treatment devices by using tissue-mimicking phantom systems (Tissue-Mimicking Material, TMM) is indispensible in pre-clinical studies. In this survey, in order to examine the viscous heating effect of thermocouples, temperature rise due to excitation of a HIFU transducer has been measured by using two different types of T-type thermocouples embedded inside the TMM. In order to obtain the temperature values from the thermocouples more effectively, a PC-based multi-channel data acquisition and monitoring interface were developed by using LabVIEW graphical development platform. Measurements were carried out inside a TMM which has characteristics like human liver tissue. Temperature rise has been characterized for the different input ultrasonic powers (25 W, 38 W and 50 W) and frequencies (1,1 MHz and 3.3 MHz) and the time intervals applied (1 s, 2 s and 3 s). The temperature values obtained from two different types of T-type thermocouple were compared to each other and discrepancies between these two different thermocouple measurements were shown with the study.
机译:通过使用组织模拟幻像系统(组织 - 模仿材料,TMM)在临床前研究中不可或缺的实现对聚焦超声处理装置的温度表征。在该调查中,为了检查热电偶的粘性加热效果,通过使用嵌入在TMM内部的两种不同类型的T型热电偶来测量由于HIFU换能器的激励引起的温度上升。为了更有效地从热电偶获得温度值,通过使用LabView图形开发平台开发了基于PC的多通道数据采集和监控接口。测量在TMM内进行,其具有人体肝组织等特征。升温的特征在于不同的输入超声波(25W,38W和50 W)和频率(1,1MHz和3.3MHz)和施加的时间间隔(1 S,2 S和3秒)。比较了从两种不同类型的T型热电偶获得的温度值彼此相互比较,并且在研究中显示了这两个不同的热电偶测量之间的差异。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号