首页> 外文会议>IEEE International Symposium on Medical Measurements and Applications >Concentration detection in water-glucose mixtures for medical applications using ultrasonic velocity measurements
【24h】

Concentration detection in water-glucose mixtures for medical applications using ultrasonic velocity measurements

机译:使用超声速度测量来检测医疗应用中的水葡萄糖混合物中的浓度

获取原文

摘要

This paper presents the possibility to measure the blood sugar content detecting different ultrasonic velocities in water-glucose mixtures by using two ultrasonic transducers. With the transmission-trough technique and a fixed transducer distance the required wave propagation time can be measured precisely. The ultrasonic actor transducer, excited at the same frequency of f = 2 MHz during different mixtures, uses a burst signal to separate reflections from the direct path. With this technique a very sensitive glucose-water concentration of less than 30 mg/dl can be separated.
机译:本文提出了使用两个超声换能器测量血糖含量的可能性,以检测水-葡萄糖混合物中不同的超声速度。利用传输槽技术和固定的传感器距离,可以精确地测量所需的波传播时间。在不同混合物中以f = 2 MHz的相同频率激发的超声演员换能器,使用突发信号来分离来自直接路径的反射。利用这种技术,可以分离出小于30 mg / dl的非常敏感的葡萄糖-水浓度。