Numerical Estimation of Peak/Average SAR Ratio for Different Thorax Models

机译:不同胸部模型的峰值/平均SAR比数值估计

获取原文

摘要

The estimation of Specific Absorption Rate (SAR) during a Magnetic Resonance Imaging (MRI) scan is an important safety procedure. In this work, a birdcage body coil model is used for radiofrequency (RF) excitation of several inhomogeneous human thorax models with different sizes and weights. To study the peak/average SAR ratio correlation with sample metrics, numerical simulations using the Finite Difference Time Domain (FDTD) method are performed to estimate the peak and average SAR values on the entire sample volume. Results for 11 different thorax models indicate a strong correlation between the peak/average SAR value and the sample mass/height ratio.
机译:在磁共振成像(MRI)扫描期间的特定吸收率(SAR)的估计是重要的安全程序。在这项工作中,鸟笼体线圈模型用于具有不同尺寸和重量的几种非均匀人胸部模型的射频(RF)激发。为了研究与采样度量的峰值/平均SAR比率相关,使用有限差分时域(FDTD)方法的数值模拟来估计整个样本体积上的峰值和平均SAR值。结果11个不同的胸部模型表示峰值/平均SAR值与样品质量/高度比之间的强关系。