摘要

Recording neural signals using implanted Pt-electrodes in a tripolar arrangement is affected by myoelectric interference present at the interface. In this paper we present a way of realizing the frequency response exhibited by a constant phase element (CPE) of metal electrodes and use this for the removal of interference. Simulation results show that our CPE is very close to the ideal case in the bandwidth we are interested in (i.e. 500 Hz to 10 kHz), while fitting past neutralization data reveals that removing interference now requires no adjustment.
机译:在三极管排列中使用植入的Pt电极记录神经信号受界面处存在的磁铁干扰的影响。在本文中,我们提出了一种实现金属电极的恒相元件(CPE)呈现的频率响应,并用它来移除干扰。仿真结果表明,我们的CPE非常接近我们对(即500 Hz至10 kHz)感兴趣的带宽中的理想情况,同时拟合过去中和数据,显示删除干扰现在不需要调整。