This work presents measurements of entrance skin dose (ESD) delivered by the imaging systems that are needed for treatment planning of arteriovenous malformations (AVM) by stereotactic radiosurgery (SRS). Measured ESD values are in agreement with measurements reported in the literature, and their magnitudes are less than the threshold ESD for the deterministic effects. This project pretends to establish a protocol for the quantity of image acquisitions for AVM treatment by SRS.
机译:该工作介绍了由立体定向放射牢(SRS)进行动脉畸形畸形(AVM)所需的成像系统所需的入口皮肤剂量(ESD)的测量。测量的ESD值与文献中报道的测量值一致,它们的大小小于用于确定性效果的阈值ESD。该项目假装为SRS建立用于AVM处理的图像采集量的协议。