Comparative analysis of spiral CT effective dose evaluation (thorax exams) on the basis of two estimation methods and techniques: with use of anthropomorphic body phantom and thermoluminescent detectors or using results of CT dose index (pTDI) measurements is presented. It is shown that estimates based on anthropometrical body phantom approximately 1,5 - 1,7 times exceed those received on the basis of CTDI measurements for multislice computer tomography without dependence on CT type. Effective dose estimates for single-slice CT are practically equal for both methods discussing.
机译:基于两种估计方法和技术的螺旋CT有效剂量评价(胸部考试)的比较分析:利用拟人体模和热敏发光探测器或使用CT剂量指数(PTDI)测量的结果。结果表明,基于人培素体幻象的估计约为1,5-1,7次超过基于CTDI测量对于多层计算机断层摄影的CTDI测量而不依赖CT型。对于两种方法讨论,单切片CT的有效剂量估计实际上是相等的。