Towards Improving the Statscan X-Ray Image Quality through Sliding-Mode Control of the C-Arm

机译:通过C型臂的滑模控制来改善Statscan X射线图像质量

获取原文

摘要

A method to improve image quality of the Statscan™ X-ray system is investigated. Transient errors in the trapezoidal motion profile of the scanning C-arm may cause mismatches between the detector and collimated beams from the X-ray source. This results in the partial degradation of image quality. The performances of two C-arm motion controllers were investigated using computer simulations and experiments: a standard PI and DMRISMC. The controllers are competitive at high sampling rate. However, the DMRISMC controller shows better performance motion profile tracking when slow sampling rate is used. There are strengths and weaknesses for both controllers and DMRISMC was expected to improve the image quality.
机译:研究了一种提高Statscan™X射线系统图像质量的方法。扫描C形臂的梯形运动曲线中的瞬态误差可能会导致探测器与来自X射线源的准直光束之间的不匹配。这导致图像质量的部分下降。使用计算机仿真和实验研究了两个C臂运动控制器的性能:标准PI和DMRISMC。控制器在高采样率下具有竞争力。但是,当使用慢速采样率时,DMRISMC控制器显示出更好的运动轨迹跟踪性能。控制器都有优点和缺点,并且DMRISMC有望改善图像质量。