Practical method to evaluate electromagnetic interference in active implantable medical devices

机译:评估有源植入式医疗设备中电磁干扰的实用方法

获取原文

摘要

It is known that some of external electromagnetic fields cause interference (EMI) to an active implantable medical device (AIMD) such as a cardiac pacemaker. Therefore it is important to establish a method to evaluate whether a particular electromagnetic environment would cause EMI to the AIMD or not. For this evaluation, the AIMD should be placed in a human body model, which is a kind of a saline tank. However, if the AIMD handles more than one chamber of the heart, signal leakage between the chambers comes up in the human body model. And it becomes difficult to monitor the output from only one chamber and to inject a simulated cardiac signal (EGM) only to the intended chamber. The author had worked out a new electrode that can reduce signal leakage of the output to less than 4 %, and injected EGM leakage to less than 3%, whereas the output leakage is 44% and the EGM leakage is 71% by another commonly-used method.
机译:众所周知,某些外部电磁场会对诸如心脏起搏器的有源可植入医疗设备(AIMD)产生干扰(EMI)。因此,重要的是建立一种评估特定电磁环境是否会对AIMD产生EMI的方法。为了进行此评估,应将AIMD放置在人体模型中,该模型是一种盐水罐。但是,如果AIMD处理心脏的多个腔室,则在人体模型中会出现腔室之间的信号泄漏。而且,仅从一个腔室监视输出并仅将模拟的心脏信号(EGM)注入目标腔室变得很困难。作者设计了一种新的电极,可以将输出的信号泄漏减少到小于4%,注入的EGM泄漏减少到小于3%,而另一种常见的电极则是输出泄漏为44%,EGM泄漏为71%。使用的方法。