Evaluation of the safety of users of active implantable medical devices (AIMD) in the working environment in terms of exposure to electromagnetic fields – practical approach to the requirements of European Directive 2013/35/EU

机译:在暴露于电磁场方面的工作环境中,在工作环境中评价 - 欧洲指令要求的实用方法 - 2013/35 /欧盟的实际方法

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

ObjectivesElectromagnetic fields (EMF) may cause malfunctions in electronic devices, in particular in active implantablemedical devices (AIMD), along with discomfort or health hazards to users. The use of AIMD by workers is increasing (especiallycardiac pacemakers, implantable cardioverter defibrillators and wearable insulin infusion pumps). Electromagneticfields may be much stronger in the working environment than applied in basic immunity tests of AIMD (based on EN 60601-1-2:2015 and EN 50527-1:2016). European Directive 2013/35/EU regarding the safety of workers exposed to EMF consideredthe AIMD users to be “workers at particular risk” who need an individual evaluation of EMF hazards. The studyaimed at evaluating the safety of users of AIMD in medical and industrial working environments exposed to EMFMaterial and MethodsNear the common sources of strong EMF applied in medical and industrial use, the “standard safety distances”(SSD) for AIMD users were evaluated (i.e., distances from the EMF source, where exposure drops below limits fromRecommendation 1999/519/EC and AIMD safety may be expected). The analysis is based on the results of measurementsof magnetic and electric field strengths near 127 typical devices, in their normal use.ResultsThe longest electric field relatedSSD was identified near dielectric sealers (up to 180 cm), and the longest magnetic field related SSD – near inductionheaters (up to 450 cm).ConclusionsElectromagnetic fields related AIMD malfunctions need to be considered up to severalmeters from EMF sources. The “individual safety distance,” that is sufficient to ensure the safety of a particular AIMD usermay be significantly different (usually shorter) from the presented SSD, but needs to be considered in the context of detailedsafety data from the AIMD manufacturer (if available). The labelling indicating the location of the area of a strong EMFincreases safety of AIMD users in the work environment. Int J Occup Med Environ Health 2018;31(6):795–808
机译:目标电元磁场(EMF)可能导致电子设备的故障,特别是在活性ImplantableDical设备(AIMD)中,以及对用户的不适或健康危害。工人使用AIMD正在增加(特别是心脏起搏器,可植入的心脏过硬纤维化剂和可穿戴胰岛素输注泵)。电磁场在工作环境中可能比应用于AIMD的基本免疫试验(基于EN 60601-2:2015和EN 50527-1:2016)。欧洲指令2013/35 /欧盟有关暴露于EMF的工人安全认为艾滋病用户认为需要对EMF危害进行个人评估的“特定风险的工人”。在评估艾滋病医疗和工业工作环境中的AIMD用户的安全性的拟计评估了在医疗和工业用途中应用的强大强大的常见来源,评估了“标准安全距离”(即标准安全距离)(即,距离EMF源的距离,其中曝光低于限制的限制,可以预期1999/519 / EC和AIMD安全性)。该分析基于127典型设备附近的磁性电场强度的测量结果,在正常使用中。近介电密封剂(高达180厘米)和最长磁场相关的SSD识别最长的电场相关的电场相关性和最长的电场诱导仪(高达450厘米)。结论电气磁场相关的AIMID故障需要被视为从EMF来源的数额。 “个人安全距离”足以确保特定瞄准型USERMay的安全性来自所呈现的SSD的大大不同(通常较短),但需要在AIMD制造商的详细信息数据上进行考虑(如果可用) 。标签,指示在工作环境中瞄准用户的强大Emincurase安全区域的位置。 INT J占MED环境2018; 31(6):795-808