Precision of Iodine Quantification in Hepatic CT: Effects of Reconstruction (FBP and MBIR) and Imaging Parameters

机译:肝脏CT中碘定量的精度:重建(FBP和MBIR)和成像参数的影响

获取原文

摘要

In hepatic CT imaging, the lesion enhancement after the injection of contrast media is of quantitative interest. However, the precision of this quantitative measurement may be dependent on the imaging techniques such as dose and reconstruction algorithm. To determine the impact of different techniques, we scanned an iodinated liver phantom with acquisition protocols of different dose levels, and reconstructed images with different algorithms (FBP and MBIR) and slice thicknesses. The contrast of lesions was quantified from the images, and its precision was calculated for each protocol separately. Results showed that precision was improved by increasing dose, increasing slice thickness, and using MBIR reconstruction. When using MBIR instead of FBP, the same precision can be achieved at 50% less dose. To our knowledge, this is the first investigation of the quantification precision in hepatic CT imaging using iterative reconstructions.
机译:在肝脏CT成像中,注射造影剂后的病变增强具有定量感兴趣。然而,这种定量测量的精度可以取决于成像技术,例如剂量和重建算法。为了确定不同技术的影响,我们扫描碘化肝脏幻像与不同剂量水平的采集协议,以及用不同算法(FBP和MBIR)和切片厚度的重建图像。病变对比度从图像量化,并且分别针对每个协议计算其精度。结果表明,通过增加剂量,增加的切片厚度和使用MBIR重建,改善了精度。使用MBIR而不是FBP时,可以以50%的剂量实现相同的精度。据我们所知,这是利用迭代重建进行肝CT成像中定量精度的第一次调查。