ECG motion artefact reduction improvements of a chest-based wireless patient monitoring system

机译:基于胸部的无线患者监护系统的ECG运动伪影减少改进

获取原文

摘要

An evaluation of motion artefact for a newly CE approved wireless bodyworn monitoring device is presented. This evaluation has shown that the system under test has greatly reduced motion artefact with comparison to an FDA-approved leaded system. Analysis of physiological data, such as quality of ECG signal, accuracy of recording of heart rate, temperature and ECG R-R interval has shown the system to offer high fidelity recordings and a robust service during a range of basic movements. Presented results have shown that the average difference in heart rate between the prototype and the reference device was 3.8bpm with standard deviation of 12.4bpm. Temperature analysis indicated the average difference between the prototype and the reference device was 5.66°C, with standard deviation of 0.44°C. R-R interval analysis highlighted mean interval difference as 78.96ms with standard deviation of 123.1ms. In general, the user activity of bending had highest errors due to the considerable torso movement.
机译:提出了一种新的CE认证的无线人体穿戴监测设备对运动伪影的评估。该评估表明,与FDA批准的引导系统相比,被测系统大大减少了运动伪影。对生理数据(例如ECG信号的质量,心率,温度和ECG R-R间隔的记录准确性)的分析表明,该系统可以在一系列基本运动中提供高保真度记录和强大的服务。提出的结果表明,原型与参考设备之间的平均心率差为3.8bpm,标准差为12.4bpm。温度分析表明,原型与参考器件之间的平均差为5.66°C,标准偏差为0.44°C。 R-R间隔分析突出显示平均间隔差异为78.96ms,标准偏差为123.1ms。通常,由于相当大的躯干运动,弯曲的用户活动具有最高的误差。