首页> 外文会议>2010 Computers in Cardiology >Blood oxygen level measurement with a chest-based Pulse Oximetry prototype system
【24h】

Blood oxygen level measurement with a chest-based Pulse Oximetry prototype system

机译:使用基于胸部的脉搏血氧仪原型系统测量血氧水平

获取原文

摘要

This paper presents a prototyped novel chest-based Pulse Oximetry system, and reports on test results from comparative trials with a commercially available finger-based Pulse Oximetry system using several human subjects. The system was iteratively optimized through adjustment of optical component alignment (angular position, component distance, photosensitive area etc.) and through fine-tuning of LED intensity and receiver sensitivity. This work is significant and timely as it provides compelling evidence that SpO2 measurements from the chest offer a genuine commercial solution for bedside and ambulatory vital-signs monitoring.
机译:本文介绍了原型的新型基于胸部的脉搏血氧饱和度测定系统,并报告了比较试验的结果,该比较试验是使用市售的基于手指的脉搏血氧饱和度测定系统进行的,该系统使用了多个人类受试者。该系统通过调整光学组件的对齐方式(角度位置,组件距离,感光面积等),以及通过对LED强度和接收器灵敏度进行微调来反复优化。这项工作意义重大且及时,因为它提供了令人信服的证据,证明胸部的SpO 2 测量值可为床旁和非卧床生命体征监测提供真正的商业解决方案。