首页> 外文会议>2010 International Conference on Machine Vision and Human-Machine Interface >Bismuth-film Glassy Carbon Electrode for the Determination of Lead by Differential Pulse Stripping Voltammetric
【24h】

Bismuth-film Glassy Carbon Electrode for the Determination of Lead by Differential Pulse Stripping Voltammetric

机译:铋膜玻碳电极微分脉冲溶出伏安法测定铅

获取原文

摘要

This paper presents a new technology using Bismuth-film to determinate lead instead of mercury film. It is proved in the experiment that, in the optimum measuring situation (when the pH is 4, deposition time is 150s, and the deposition potential is 0.8.), the lead concentration and stripping peak current shows a good linear relationship in the measuring range of 10 to 110ug/L, which is I=0.1002C+0.0539. The experimental analysis shows that the concentration of lead in tea samples is 1.667μg / g, and the recovery rate is 91.2% ~ 101.8%. It is proved that the method of bismuth film electrode is used to measure lead with advantages of convenience, speediness, sensitivity and accuracy, etc.
机译:本文介绍了一种使用铋膜代替汞膜测定铅的新技术。实验证明,在最佳测量条件下(pH为4时,沉积时间为150s,沉积电势为0.8。),在测量范围内,铅浓度与溶出峰电流呈良好的线性关系。 10至110ug / L,即I = 0.1002C + 0.0539。实验分析表明,茶叶样品中铅的含量为1.667μg/ g,回收率为91.2%〜101.8%。实践证明,铋膜电极法用于铅的测定具有方便,快捷,灵敏,准确等优点。