首页> 外文会议>Conference on physics of medical imaging >A Simple Method for Evaluating Image Quality of Screen-Film System Using A High-Performance Digital Camera
【24h】

A Simple Method for Evaluating Image Quality of Screen-Film System Using A High-Performance Digital Camera

机译:使用高性能数码相机评估屏幕胶片系统图像质量的简单方法

获取原文

摘要

Screen-film systems are used in mammography even now. Therefore, it is important to measure their physical properties such as modulation transfer function (MTF) or noise power spectrum (NPS). The MTF and NPS of screen-film systems are mostly measured by using a microdensitometer. However, since microdensitometers are not commonly used in general hospitals, it is difficult to carry out these measurements regularly. In the past, Ichikawa et al. have measured and evaluated the physical properties of medical liquid crystal displays by using a high-performance digital camera. By this method, the physical properties of screen-film systems can be measured easily without using a microdensitometer. Therefore, we have proposed a simple method for measuring the MTF and NPS of screen-film systems by using a high-performance digital camera. The proposed method is based on the edge method (for evaluating MTF) and the one-dimensional fast Fourier transform (FFT) method (for evaluating NPS), respectively. As a result, the MTF and NPS evaluated by using the high-performance digital camera approximately corresponded with those evaluated by using a microdensitometer. It is possible to substitute the calculation of MTF and NPS by using a high-performance digital camera for that by using a microdensitometer. Further, this method also simplifies the evaluation of the physical properties of screen-film systems.
机译:屏幕胶片系统甚至在乳腺摄影中也被使用。因此,重要的是测量它们的物理特性,例如调制传递函数(MTF)或噪声功率谱(NPS)。屏幕胶片系统的MTF和NPS大多通过使用微密度计进行测量。但是,由于微密度计在普通医院中并不常用,因此很难定期进行这些测量。过去,Ichikawa等人。已通过使用高性能数码相机测量和评估了医用液晶显示器的物理性能。通过这种方法,可以在不使用微密度计的情况下轻松地测量屏幕胶片系统的物理性能。因此,我们提出了一种通过使用高性能数码相机来测量屏幕胶卷系统的MTF和NPS的简单方法。所提出的方法分别基于边缘方法(用于评估MTF)和一维快速傅里叶变换(FFT)方法(用于评估NPS)。结果,使用高性能数码相机评估的MTF和NPS与使用微密度计评估的MTF和NPS大致相对应。通过使用高性能数码相机,可以用微密度计代替MTF和NPS的计算。此外,该方法还简化了对屏幕胶片系统物理性能的评估。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号