首页> 外文会议>Frontiers in Biomedical Devices Conference 2006 >REAL-TIME GLUCOSE SENSING USING A MEMS-BASED BI-MATERIAL DEFLECTING MEMBRANE
【24h】

REAL-TIME GLUCOSE SENSING USING A MEMS-BASED BI-MATERIAL DEFLECTING MEMBRANE

机译:使用基于MEMS的双材料偏转膜进行实时葡萄糖感测

获取原文

摘要

A HEMA polymer, which swells in the presence of glucose, can be used in a laminated membrane and incorporated into a MEMS device to measure, in real-time, the concentration of glucose in solution. Design of such a device requires careful experiments to determine the swelling behavior of the HEMA, as well as the composite response of a laminated system. We have shown that the swelling can be controlled by changing the cross-link density. Incorporation of the polymer into a device requires a novel fabrication approach utilizing XeF_2 vapor phase etching. Current experiments are aimed at establishing a theoretical model to predict the mechanical properties of a HEMA/PDMS bi-layer system.
机译:在葡萄糖存在下会膨胀的HEMA聚合物可用于层压膜中,并结合到MEMS器件中,以实时测量溶液中葡萄糖的浓度。这种设备的设计需要仔细的实验​​以确定HEMA的溶胀行为以及层压系统的复合响应。我们已经表明,膨胀可以通过改变交联密度来控制。将聚合物掺入到器件中需要一种利用XeF_2气相蚀刻的新颖制造方法。当前的实验旨在建立理论模型以预测HEMA / PDMS双层系统的机械性能。