The paper presents a numerical, worst-case study of the coupling between coagulators and implanted neurostimulation systems under consideration of the human body. For this purpose, the devices regarded were characterized and modeled and a special human body model has been implemented. With this model, whose electrical properties correspond to a weighted average of those of different tissues of the human body, the influence of the latter can be taken into account in the computations. Different parameters, such as position of the neurostimulation system and its electrode, coagulation frequency and coagulation point on the surface of the human body, are considered to achieve worst-case results. The investigation shows that, for the scenario described, strong peaks occur in the resulting graphs arising from resonance effects on the coagulator cable.
机译:本文介绍了在考虑人体因素的情况下,对凝结器和植入的神经刺激系统之间的耦合进行的最坏情况的数值研究。为此,对所考虑的设备进行了表征和建模,并实现了特殊的人体模型。使用该模型,其电特性对应于人体不同组织的电特性的加权平均值,因此可以在计算中考虑后者的影响。人们认为,诸如神经刺激系统及其电极的位置,人体表面的凝结频率和凝结点之类的不同参数可以达到最坏的情况。研究表明,对于所描述的情况,由于对凝结器电缆的共振效应而导致的结果图中出现了很强的峰。