Analytical Noise and Slice-sensitivity-profile Models for Multi-slice Helical CT

机译:多层螺旋CT的分析噪声和切片灵敏度曲线模型

获取原文

摘要

Multi-slice computed tomography (MSCT) is one of the most recent technological advancements in CT. Several experimental studies have been conducted in the last few years on the performance of MSCT. However, there is a lack of theoretical analysis on the slice sensitivity profile (SSP) and noise performance. In this paper, we derive several clpsed-form expressions to characterize these performance parameters under different detector configurations and acquisition modes. Following the common practice, the expressions are explicitly described for regions near the iso-center, although the same approach can be used to describe system performances away from the iso-center. Our models are validated against phantom experiments.
机译:多层计算机断层扫描(MSCT)是CT的最新技术进步之一。在过去的几年中,已经对MSCT的性能进行了一些实验研究。但是,缺乏对切片灵敏度曲线(SSP)和噪声性能的理论分析。在本文中,我们导出了几种形式的表达式,以表征不同检测器配置和采集模式下的这些性能参数。遵循惯例,虽然可以使用相同的方法来描述远离等中心点的系统性能,但是明确表示了等中心点附近区域的表达式。我们的模型已针对幻像实验进行了验证。