Spectral band calibration of the Total Ozone Mappling Spectrometer (TOMS) using a tunable laser technique

机译:使用可调谐激光技术对总臭氧谱图仪(TOMS)进行光谱带校准

获取原文

摘要

The Total Ozone Mapping Spectrometer (TOMS) provides daily global mapping of the total column ozone in the earth's atmosphere. It does this by measuring the solar irradiance and the backscattered solar radiance in 6 spectral bands falling within the range from 308.6 nm to 360 nm. THe accuracy of the ozone retrieval is highly dependent on the knowledge of the transfer characteristics and center wavelength for each spectral band. A 0.1 nm wavelength error translates to a 1.6percent error in ozone. Several techniques have historically been used to perform the wavelength calibration of the TOMS instruments. These methods include the use of film and reference spectra from low-pressure spect ral lone lamps and the use of continuum sources with a narrow-band scanning mochromater.
机译:臭氧总谱仪(TOMS)每天提供地球大气中总臭氧柱的全球总图。它通过测量在308.6 nm至360 nm范围内的6个光谱带中的太阳辐照度和反向散射太阳辐照度来实现此目的。臭氧回收的准确性高度依赖于每个光谱带的传输特性和中心波长的知识。波长误差为0.1 nm会导致臭氧误差为1.6%。历史上已经使用了几种技术来执行TOMS仪器的波长校准。这些方法包括使用来自低压光谱孤灯的胶片和参考光谱,以及使用具有窄带扫描复色仪的连续谱源。