We present a new multipurpose phantom for routine dosimetric and image quality assurance (QA) testing of a tomotherapy unit. The phantom is designed for verification of delivered dose from the intensity modulated radiation therapy (IMRT) fan beams used to provide conformal dose distributions. The phantom is used to assess imaging characteristics and verify CT number calibration for the integrated megavoltage computed tomography (MVCT) detector mounted on the gantry across from the therapy linac. Functional requirements are outlined for the unique dosimetric and imaging needs of the QA tests of the tomotherapy system. The 36 cm long phantom is constructed mostly of Solid Water/sup TM/ and consists of 6 adjacent discs. The phantom consists of head and torso modules. The phantom is divided into imaging and dosimetric regions. For fan beam IMRT dosimetry, both a standard Farmer-style ionization chamber and a CT pencil chamber are used to measure the dose and dose rate under a variety of delivery conditions and phantom heterogeneities. The feasibility of using an existing kV CT chamber at therapy energies was investigated. The response to irradiation by small individual fields (7.6 mm) along its length is within one standard deviation of the average for the central 80% of its active volume (10 cm). For imaging QA, this phantom contains cavities to accommodate standard inserts to test spatial resolution, low contrast detectability, noise and edge contrast. Addition inserts are required for helical pitch verification and detector centering.
机译:我们提出了一种新的多功能体模,用于对tomotherapy装置进行常规剂量和图像质量保证(QA)测试。该体模设计用于验证来自强度调制放射疗法(IMRT)扇形光束提供的剂量,该剂量用于提供保形剂量分布。该体模用于评估成像特性并验证安装在治疗直线加速器对面的龙门架上的集成兆伏计算机断层扫描(MVCT)检测器的CT数量校准。概述了功能要求,以满足断层扫描系统QA测试的独特剂量和成像需求。 36厘米长的幻像主要由Solid Water / sup TM /构成,由6个相邻的圆盘组成。幻影由头部和躯干模块组成。体模分为成像区和剂量区。对于扇形束IMRT剂量测定,标准的Farmer型电离室和CT笔室均用于在各种输送条件和幻影异质性下测量剂量和剂量率。研究了在治疗能量下使用现有的kV CT腔室的可行性。沿其长度方向的小单个场(7.6 mm)对辐照的响应在其活动体积的中心80%(10 cm)的平均值的平均值的一个标准偏差内。对于QA成像,该模型包含腔体,以容纳标准插入物以测试空间分辨率,低对比度可检测性,噪声和边缘对比度。需要附加插件来进行螺旋螺距验证和检测器对中。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号