Effects of radiation dose and display contrast on low-contrast phantom image visibility

机译:辐射剂量和显示器对比度对低对比度幻像图像可见性的影响

获取原文

摘要

Abstract: Computed radiography (CR) radiographs were generated of a low contrast phantom with 5 mm diameter disks. The radiation exposure incident on the imaging plate was varied from approximately 0.1 mR to approximately 10 mR, with the phantom images printed to film using a range of display contrast settings. Changing the radiation exposure by two orders of magnitude had only a modest effect on disk detection performance (approximately 20%), and much less than predicted by signal detection theory for the perception of noise limited objects. For images generated at approximately 1 and approximately 10 mR, increasing the display contrast markedly improved the disk detection performance (approximately 50%). There was approximate agreement between the experimental data and the corresponding theoretical predictions for the detection of contrast limited objects. For the contrast detail phantom employed in this study, disk detection was primarily contrast limited, with image noise being relatively unimportant. Lesion detection with an anthropomorphic phantom containing a structured background wold be unlikely to change this conclusion, since noise is expected to be most important for low contrast objects viewed against a uniform background. Contrast enhancement, as opposed to increasing radiation exposure, is therefore the method of choice for improving the detection of 5 mm diameter sized low contrast lesions in CR images. !14
机译:摘要:用直径为5 mm的圆盘的低对比度体模生成了计算机射线照相(CR)射线照片。入射在成像板上的辐射曝光量从约0.1 mR到约10 mR变化,并且使用一系列显示对比度设置将幻影图像打印到胶片上。将辐射暴露量更改两个数量级对磁盘检测性能仅产生了中等程度的影响(大约20%),远小于信号检测理论预测的噪声受限对象的感知。对于以大约1 mR和大约10 mR生成的图像,提高显示对比度可以显着提高磁盘检测性能(大约50%)。实验数据和相应的理论预测之间在对比度受限物体的检测上有近似的一致性。对于本研究中使用的对比度细节模型,磁盘检测主要受对比度限制,图像噪声相对不重要。用拟人化体模包含结构化背景阴影的病灶检测不太可能改变这一结论,因为对于在统一背景下观察到的低对比度对象,噪声被认为是最重要的。因此,与增加辐射暴露相反,对比度增强是用于改善对CR图像中5毫米直径大小的低对比度病变的检测的选择方法。 !14

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号