首页> 外文会议>Symposium on recent advance in the chemistry and physics of fullerenes and related materials;Meeting of The Electrochemical Society >The QCM/HCC: simultaneous, isothermal, high sensitivity measurements of mass change and heat flow in polymer and fullerene films
【24h】

The QCM/HCC: simultaneous, isothermal, high sensitivity measurements of mass change and heat flow in polymer and fullerene films

机译:QCM / HCC:同时,等温,高灵敏度地测量聚合物和富勒烯薄膜中的质量变化和热流

获取原文

摘要

A new, high sensitivity method has been developed of measuring simultaneously the mass deposited on a surface and the resulting heat flow to that surface. The method employs a quartz crystal microbalance (QCM) coupled thermally to a heat conduction calorimeter (HCC), and the combination can measure mass changes to +-0.02 mu g and heat flows to +-3 mu W. The construction of the apparatus and its performance are described. Measurements on the sorption of chloroform vapor by an aliphatic polyurethane film 3 mu m thick and on the sorption of bromobenzene vapor by a C_(60) film 50 nm thick are presented.
机译:已经开发出一种新的高灵敏度方法,该方法可以同时测量沉积在表面上的质量以及由此产生的流向该表面的热量。该方法采用热耦合至导热量热仪(HCC)的石英晶体微天平(QCM),该组合可以测量质量变化为+ -0.02μg,热流为+ -3μW。描述其性能。提出了对厚度为3μm的脂肪族聚氨酯膜对氯仿蒸气的吸附以及对厚度为50 nm的C_(60)膜对溴苯蒸气的吸附的测量方法。