The paper presents a 3D colour ultrasound imaging system for the estimation of flow parameters downstream of prosthetic heart valves. The system consists of an ATL Ultramark 9 HDI colour Doppler ultrasound machine with a 38 mm aperture and a 5 MHz high resolution linear array transducer, a 3D ultrasound imaging system, and an in vitro circulation system under which the prosthetic heart valves are tested. Preliminary results obtained in the case of St. Jude prosthetic heart valves show that the technique can be used successfully to estimate the mean flow velocity, the cross-sectional area of the forward flow, and the impact of the jets. Further work is required for more quantitative evaluation of flow velocity. In this context, some suggestions are presented in order to improve the velocity estimation.
机译:本文提出了一种3D彩色超声成像系统,用于估计人工心脏瓣膜下游的血流参数。该系统由具有38 mm孔径的ATL Ultramark 9 HDI彩色多普勒超声仪和5 MHz高分辨率线性阵列换能器,3D超声成像系统以及体外循环系统组成,在该系统下测试人造心脏瓣膜。在St. Jude人工心脏瓣膜的情况下获得的初步结果表明,该技术可以成功地用于估计平均流速,正向流动的横截面积和射流的影响。需要进一步的工作以更定量地评估流速。在这种情况下,提出了一些建议以改善速度估计。