首页> 外文会议>Medical Imaging 1997: Ultrasonic Transducer Engineering >Parallel beam-forming architectures: design and simulation
【24h】

Parallel beam-forming architectures: design and simulation

机译:并行波束形成架构:设计和仿真

获取原文

摘要

Abstract: A model of a medical ultrasound imaging system can be broken into four major subsystems: the transducer, the array, the beamformer, and display system. In this work, each sub- system is modeled separately and combined to simulate the complete system. Transducer elements are modeled as three port systems with two mechanical ports and one electrical port. The array is modeled as a product of a point source array and finite radiating elements. The beamforming system is modeled as a frequency domain beamformer with depth focusing, utilizing a parallel computing architecture. Each of these systems offers flexibility to model a one dimensional linear phased array, a 1.5D array or a two dimensional array, or an array of arbitrary shape. !12
机译:摘要:医学超声成像系统的模型可以分为四个主要子系统:换能器,阵列,波束形成器和显示系统。在这项工作中,每个子系统分别建模并组合以模拟整个系统。换能器元件被建模为具有两个机械端口和一个电气端口的三端口系统。将该阵列建模为点源阵列和有限辐射元素的乘积。利用并行计算架构,将波束成形系统建模为具有深度聚焦功能的频域波束成形器。这些系统中的每一个都提供了对一维线性相控阵列,1.5D阵列或二维阵列或任意形状的阵列建模的灵活性。 !12