MEASUREMENT ACCURACY OF ULTRASOUND HEART RATE MONITOR

机译:超声心率监测仪的测量精度

获取原文

摘要

In order to ensure the accuracy of the fetal heart rate (FHR) obtained from the ultrasound Doppler data, we measure the fetal electrocardiogram (ECG) directly and obtain the Doppler data simultaneously. The FHR differences of the Doppler data from the direct ECG data are considerably small, concentrated at 0 bpm (beats per minute), and are practically symmetrical. The distribution was found to be close to the Student's t distribution. Furthermore, the 1/f fluctuation is observed both with the ultrasound Doppler FHR and with the direct ECG FHR. Thus, it is confirmed that the FHR observation and observation of the 1/f fluctuation using the ultrasound Doppler FHR are as useful as the direct ECG.
机译:为了确保从超声多普勒数据获得的胎儿心率(FHR)的准确性,我们直接测量胎儿心电图(ECG)并同时获得多普勒数据。从直接ECG数据的多普勒数据的FHR差异相当小,浓缩,在0bpm(每分钟节拍),并且实际上是对称的。发现分布接近学生的T分配。此外,通过超声波多普勒FHR和直接ECG FHR观察到1 / F波动。因此,证实使用超声多普勒FHR的FHR观察和观察使用超声波多普勒FHR的波动与直接ECG一样有用。