首页> 外文会议>Frequency Control Symposium, 1997., Proceedings of the 1997 IEEE International >An acoustic plate mode study of complementary and non-complementary antigen/antibody reactions
【24h】

An acoustic plate mode study of complementary and non-complementary antigen/antibody reactions

机译:互补和非互补抗原/抗体反应的声板模式研究

获取原文

摘要

Acoustic plate mode sensors have been used to monitor immunochemical reactions as a function of antigen concentration. In the studies, antibodies were covalently linked to the gold-coated sensing surface via mercaptoethanol, aminosilane and glutaraldehyde. Two antigen/antibody model systems have been used which differ in their ability to mutually bind one another. For sensor operation at about 150 MHZ a detection limit of about 0.5 /spl mu/g/ml was obtained in both cases. No significant difference between the two systems was found for the value of the binding constants. They amount to about 1.10/sup 8/ 1/mol and fall well into the range of binding constants reported for homogeneous immunoassays. A comparison of the sensor response obtained for the two model systems shows that about 70% of the immobilized antibodies are active.
机译:声学板模式传感器已被用来监测免疫化学反应作为抗原浓度的函数。在研究中,抗体通过巯基乙醇,氨基硅烷和戊二醛共价连接到涂金的传感表面上。已经使用了两种抗原/抗体模型系统,它们彼此结合的能力不同。对于在约150 MHZ的传感器操作,在两种情况下均获得约0.5 / spl mu / g / ml的检测限。结合常数的值在两个系统之间没有发现显着差异。它们总计约1.10 / sup 8/1 // mol,并且完全落入均相免疫测定所报道的结合常数的范围内。对两种模型系统获得的传感器响应的比较表明,约70%的固定抗体具有活性。