首页> 外文会议>Ophthalmic Technologies V >Reproducibility study of placido-based corneal topography systems
【24h】

Reproducibility study of placido-based corneal topography systems

机译:基于普拉西多的角膜地形图系统的再现性研究

获取原文

摘要

Abstract: We have developed a computer software program that analyzes the reproducibility of corneal topography systems. Using a Macintosh computer with Microsoft QuickBasic programming we created a program that records and analyzes the corneal power of 72,000 points per topography map (200 points over each one degree meridian) by linear interpolation of the Placido-based ring data. Reproducibility studies on 7 normal and 7 abnormal (pathological and postsurgical) corneas were performed with 2 Placido ring based topography systems with EyeSys Corneal Analysis System and the Humphrey MasterView system. Five exams were performed by two technicians each. The mean, range and standard deviation of corneal power was calculated for all points on a map, and represented pictorially for each patient. Statistical analysis of standard deviation at each point was used in assessing the reproducibility between systems and between technicians. The topographic maps of mean power range and standard deviation are presented for selected patients. The overall mean standard deviation of power and the means within concentric annular zones and quadrants are presented for each system and technician for both normal and abnormal cornea. Intraobserver variability of each system were not significantly different for both normal and abnormal corneas. There was no significant difference between technicians, nor were there regional differences in variability.standard deviation of power is presented for each system and technician comparing intersystem and intraobserver variability for both normal and abnormal corneas. A precise method of analyzing multiple variation of reproducibility in corneal topography is now available through a new proprietary software system.!14
机译:摘要:我们开发了一种计算机软件程序,用于分析角膜地形图系统的可重复性。使用带有Microsoft QuickBasic编程的Macintosh计算机,我们创建了一个程序,该程序通过基于Placido的环数据的线性插值来记录和分析每个地形图的72,000点(每个一度子午线200点)的角膜屈光力。使用2个基于Placido环的地形系统,​​EyeSys角膜分析系统和Humphrey MasterView系统,对7个正常和7个异常(病理和术后)角膜进行了再现性研究。两名技术人员各进行了五次考试。计算地图上所有点的角膜屈光力的平均值,范围和标准偏差,并以图形方式表示每个患者。每个点的标准偏差的统计分析用于评估系统之间和技术人员之间的可重复性。显示了所选患者的平均功率范围和标准偏差的地形图。为正常和异常角膜的每个系统和技术人员提供了功率的平均平均标准偏差以及同心环形区域和象限内的均值。正常和异常角膜的每个系统的观察者内变异性均无显着差异。技术人员之间没有显着差异,变异性方面也没有区域差异。针对每个系统提供了标准功率偏差,技术人员比较了正常角膜和异常角膜的系统间和观察者内部变异性。现在,通过新的专有软件系统,可以找到一种精确的方法来分析角膜地形图的可重复性的多种变化!14

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号