首页> 外文会议>International Conference on Neutrons and Their Applications >Problems in determining the absorbed dose in thermal neutron and gamma ray mixed fields: a tissue-equivalent phantom dosimeter for boron neutron capture therapy (BNCT)
【24h】

Problems in determining the absorbed dose in thermal neutron and gamma ray mixed fields: a tissue-equivalent phantom dosimeter for boron neutron capture therapy (BNCT)

机译:测定热中子和γ射线混合田中吸收剂量的问题:硼中子捕获治疗的组织当量模拟剂量计(BNCT)

获取原文

摘要

In boron neutron capture therapy (BNCT) the absorbed dose consists of the contribution of many radiation components, and its spatial distribution is highly nonuniform. Three- dimensional dose determinations are therefore to be done, and the utilized dosimeters have to perturb the radiation field as little as possible. Thermoluminescent dosimeters (TLDs) are widely utilized for mapping the radiation field. But TLDs with high sensitivity for thermal neutrons have shown an irreversible damage from such neutrons producing an alteration in the response. A tissue equivalent phantom dosimeter is proposed allowing a 3-D determination of absorbed dose by an NMR imaging of the phantom.
机译:在硼中子捕获疗法(BNCT)中,吸收剂量由许多辐射组分的贡献组成,其空间分布高度不均匀。因此要完成三维剂量测定,并且使用的剂量计必须尽可能少地扰乱辐射场。热辐发光剂量计(TLD)被广泛用于映射辐射场。但热井高灵敏度的TLD显示了从这些中子产生反应改变的不可逆损坏。提出了一种组织当量模瞳剂剂量表,允许通过脉体的NMR成像进行3-D测定吸收剂量。

著录项