The authors discuss the theoretical aspects of the extension of the calibration techniques for network analyzers to the measurement of nonlinear devices. An examination of these theoretical aspects makes it possible to pinpoint the fundamental calibration problems that must be solved before the performance of nonlinear device measurements is comparable with the achieved performance of linear system measurements. It is shown that the calibration procedure for measuring nonlinear one-port devices with low- and high-frequency network analyzers can be reduced to a full linear calibration procedure and determination of a complex scaling factor. This scaling factor can be determined in two steps: amplitude and phase calibration. For the amplitude calibration, the use of a power meter, traceable to a standard, is sufficient. For the phase calibration, a phase reference is needed. This can be a known active device or a reference generator.
机译:作者讨论了将网络分析仪的校准技术扩展到非线性设备测量的理论方面。对这些理论方面的研究使得可以查明在非线性设备测量的性能与线性系统测量的性能可比之前必须解决的基本校准问题。结果表明,使用低频和高频网络分析仪测量非线性单端口设备的校准过程可以简化为完整的线性校准过程,并且可以确定复杂的比例因子。该缩放因子可以通过两个步骤确定:幅度和相位校准。对于幅度校准,使用可溯源至标准的功率计就足够了。对于相位校准,需要一个相位参考。这可以是已知的有源设备或参考生成器。