The authors introduce a simple method by which the robust calibration technique developed by S.K. Judah (1985) may be unified with the thru-reflect-line technique, yielding initial estimates of the six-port instrumental parameters that are insensitive to noise and random errors. This calibration technique has been employed in a single six-port reflectometer operating at 2 GHz. Experimental experience has shown the technique to be robust; problems encountered using the method recommended by G.F. Engen (1979) have not arisen. Investigation by computer simulation has also shown the technique to be reliable. This methodology requires a substantial increase in software effort owing to the effective operational requirement of two full calibration techniques.
机译:作者介绍一种简单的方法,由S.K.开发的鲁棒校准技术。 Judah(1985)可以通过直通反射线技术进行统一,从而得出对噪声和随机误差不敏感的六端口仪器参数的初始估计值。此校准技术已在工作于2 GHz的单个六端口反射仪中采用。实验经验表明该技术是可靠的。使用G.F.推荐的方法遇到的问题Engen(1979)尚未出现。通过计算机仿真的研究也表明该技术是可靠的。由于两种完全校准技术的有效运行要求,因此该方法需要大量的软件工作。