MRI data acquisition and image reconstruction from multiple, non-interacting receiver coils

机译:多个非相互作用的接收线圈进行MRI数据采集和图像重建

获取原文

摘要

A previously published acquisition and reconstruction method for magnetic resonance imaging (MRI) which allows for a reduction in the number of phase encoding acquisitions in an imaging sequence without reducing either resolution or field of view is described. Two (or more) noninteracting receiver coils simultaneously detect signals from the same tissue, and data from each of the receiver channels are separately recorded. An imaging sequence uses phase encoded echoes in the usual sense, but the reconstruction effectively calculates multiple Fourier projections of the tissue from one echo. An implementation with two receiver coils can reduce the number of needed phase encoded acquisitions by 50%. The reconstruction algorithm is analyzed, the signal-to-noise ratio is calculated, and artifacts are discussed. The technique has not yet been implementated.
机译:描述了先前公开的用于磁共振成像(MRI)的采集和重建方法,该方法允许减少成像序列中的相位编码采集的数量而不降低分辨率或视野。两个(或更多个)非相互作用的接收器线圈同时检测来自同一组织的信号,并且分别记录每个接收器通道的数据。成像序列通常使用相位编码的回波,但是重建有效地从一个回波计算出组织的多个傅立叶投影。具有两个接收器线圈的实施方案可以将所需的相位编码采集数量减少50%。分析了重建算法,计算了信噪比,并讨论了伪影。该技术尚未实现。