A new mechanical prosthetic heart valve is presented: the central-axis valve. This device was tested in vitro and compared with four other common prosthetic cardiac valves. All valves studied have the same inner orifice diameter of 22 mm. The prostheses were installed inside a transparent mitral test chamber that permits pressure-drop measurements and other flow characteristics under steady flow conditions using a blood analog fluid. The tested valves were: (a) Starr-Edwards 6120 valve, (b) Bjork-Shiley monostrut valve, (c) Medtronic-Hall valve, (d) Saint Jude Medical valve, and (e) central-axis valve. Initial tests under pulsatile flow conditions demonstrated that the new valve prototype is reliable and functions well.
机译:提出了一种新的机械人工心脏瓣膜:中心轴瓣膜。该装置在体外进行了测试,并与其他四个常见的人工心脏瓣膜进行了比较。所有研究的阀门都具有相同的22 mm内孔直径。假体安装在透明的二尖瓣测试腔内,该腔允许使用血液模拟液在稳定流量条件下进行压降测量和其他流量特性。测试的阀门为:(a)Starr-Edwards 6120阀门,(b)Bjork-Shiley单支柱阀门,(c)Medtronic-Hall阀门,(d)Saint Jude Medical阀门和(e)中心轴阀门。在脉动流量条件下的初步测试表明,新的阀门原型可靠且功能良好。