NMAR3: Normalized Metal Artifact Reduction for Cone Beam Computed Tomography

机译:NMAR3:锥形束计算机断层扫描的归一化金属伪像减少

获取原文

摘要

Metal Artifact Reduction (MAR) plays an important role in Computed Tomography (CT) research and application because severe artifacts degrade the image quality and diagnosis value if metal objects are present in the field of measurement. Although there are already many works for MAR, these works are for fan beam CT, not for cone beam CT, which is the trend and receiving much research attention. In this paper, we extend the Normalized Metal Artifact Reduction (NMAR) for fan beam CT to NMAR3 for cone beam CT, by replacing the linear interpolation in the NMAR with bi-linear interpolation. Experiments are carried out on 17 sets of spine phantom CT. 15 of them have reference CT as ground truth and 2 ones not. Both quantitative and qualitative results verified that NMAR3 outperforms the baseline method, i.e., bi-linear interpolation based method.
机译:减少金属伪影(MAR)在计算机断层扫描(CT)的研究和应用中起着重要作用,因为如果在测量领域中存在金属物体,则严重的伪影会降低图像质量和诊断价值。尽管MAR的工作已经很多,但这些工作是用于扇形束CT,而不是用于锥形束CT,这是趋势,受到了很多研究关注。在本文中,我们通过用双线性插值代替NMAR中的线性插值,将扇形束CT的归一化金属伪影减少(NMAR)扩展为锥形束CT的NMAR3。实验在17套脊柱体模CT上进行。他们中有15人将CT视为基本事实,而2人则没有。定性和定量结果均证明NMAR3优于基线方法,即基于双线性插值的方法。