Development of an integrated wristband for physical activity and pulse measurement

机译:开发用于运动和脉搏测量的集成腕带

获取原文

摘要

For measuring human daily physical activity and exercise in real time, a wristband was developed in this study, which can measure physical activity and exercise pulse at the same time together. The Wristband can provide data in three aspects: (1)PA subsystem for measuring physical activity; (2) EP subsystem for measuring Exercise Pulse; (3) HF subsystem for measuring health fitness and instruction feedback. By comparing the Counts Per Minute data of the subjects with the actigraph gt3x & tri-axial accelerometer, the difference was less than 5%; the pulse data measured by wristband and polar heart rate band were within −2.12 ∼ +2.03 beat per minute and the relative error is between −1.65% ∼ +1.39% under different exercise types and different speed intensities. Conclusion: the measurement accuracy of wristband can meet the requirements of exercise pulse measuring, and it can satisfy measurement and evaluation of physical activity.
机译:为了实时测量人的日常体育锻炼和运动,本研究开发了一种腕带,它可以同时测量体育锻炼和运动脉搏。腕带可以提供三个方面的数据:(1)用于测量身体活动的PA子系统; (2)用于测量运动脉搏的EP子系统; (3)HF子系统,用于测量健康状况和指示反馈。通过将受试者的每分钟计数数据与gt3x和三轴加速度计进行比较,差异小于5%;在不同的运动方式和速度强度下,腕带和极地心率带测量的脉搏数据在每分钟-2.12〜+ 2.03次搏动范围内,相对误差在-1.65%〜+ 1.39%之间。结论:腕带的测量精度可以满足运动脉搏测量的要求,可以满足体育锻炼的测量和评价。