Current distribution analyses of figure-eight coil for an intergrated transcutaneous energy and information transmission system for a totally-implantable artificial heart

机译:完全植入式人造心脏集成式经皮能量和信息传输系统的八字形线圈的电流分布分析

获取原文

摘要

This study describes the transmission characteristics in the case of an implantable transcutaneous energy and information transmission system using a dedicated figure-eight coil for information transmission in instances when misalignments occur in the system’s transformer. Use of high-frequency information transmission was advantageous compared to low-frequency transmissions when the figure-eight coil was used when misalignments occurred. The associated mechanism was clarified by electromagnetic field analyses. Simulations showed that at low frequencies, current flowed along the coil tracks, whereas at high frequencies, current flowed along the circular shape of the outer peripheral coil.
机译:这项研究描述了在植入式经皮能量和信息传输系统中的传输特性,该系统使用专用的八字形线圈进行信息传输,以防系统的变压器发生未对准情况。当出现未对准时使用八字形线圈时,与低频传输相比,使用高频信息传输更具优势。通过电磁场分析阐明了相关的机理。仿真表明,在低频下,电流沿线圈轨道流动,而在高频下,电流沿外围线圈的圆形流动。