首页> 外文会议>IEEE International Midwest Symposium on Circuits and Systems >Design of Multi-Channel Monopolar Biphasic Stimulator for Implantable Biomedical Applications
【24h】

Design of Multi-Channel Monopolar Biphasic Stimulator for Implantable Biomedical Applications

机译:用于植入式生物医学应用的多通道单极双相刺激器的设计

获取原文

摘要

Neuro-stimulators have been widely used in neural disorder control. This work presents a four-channel stimulator with adjustable positive and negative stimulus voltage from ±0.5V to ±8V. In the proposed stimulator, circuit operation with negative voltage has been successfully realized in a 0.25-μm 2.5V/5-V/12-V CMOS process with the grounded p-type substrate, and without suffering the p-n junction breakdown or electrical overstress issues. The proposed stimulator has been also practically verified with in-vivo animal test. By integrating the stimulator circuit with other circuit blocks together, the close-loop neuromodulation treatment can be implemented by an implantable system-on-chip (SoC) with low-power consumption.
机译:神经刺激剂已被广泛用于神经疾病的控制。这项工作提出了一种四通道刺激器,其正和负刺激电压可调范围为±0.5V至±8V。在所提出的刺激器中,在具有接地p型衬底的0.25-μm2.5V / 5-V / 12-V CMOS工艺中,已经成功实现了负电压电路操作,而没有遭受pn结击穿或电气过应力问题。所提议的刺激器也已经在体内动物试验中得到了验证。通过将刺激器电路与其他电路块集成在一起,可以通过低功耗的可植入系统级芯片(SoC)来实现闭环神经调制治疗。