首页> 外文会议>IEEE International Midwest Symposium on Circuits and Systems >H-Field Probe with Active Balun for MRI-Conditional Testing of Medical Devices
【24h】

H-Field Probe with Active Balun for MRI-Conditional Testing of Medical Devices

机译:带有有源巴伦的H场探头用于医疗设备的MRI条件测试

获取原文

摘要

This work presents the design and construction of a H-field probe with active balun for MRI (magnetic resonance imaging) conditional testing of medical devices. The H-field probe was designed using a small loop antenna with 2 mm radius to have high spatial resolution and sensitivity. It was tuned and matched at a center frequency of 128 MHz, which corresponds to 3 T MRI systems or equivalent RF exposure systems. An active balun with a gain 18.28 dB and noise figure of 0.5 dB was employed to boost the detected H-field signal level. The designed Hfield probe and active balun have been fabricated on a doublesided printed circuit board (PCB), FR4 thickness of 1.57 mm and a copper thickness of 35 μm, with overall footprint of 22 mm × 11 mm.
机译:这项工作介绍了带有有源巴伦的H场探头的设计和构造,该探头用于医疗设备的MRI(磁共振成像)条件测试。 H场探头是使用半径为2 mm的小环形天线设计的,具有很高的空间分辨率和灵敏度。它以128 MHz的中心频率进行了调谐和匹配,该中心频率对应于3 T MRI系统或等效的RF曝光系统。使用增益为18.28 dB,噪声系数为0.5 dB的有源巴伦来增强检测到的H场信号电平。设计的Hfield探头和有源巴伦已在双面印刷电路板(PCB)上制造,FR4厚度为1.57 mm,铜厚度为35μm,总占地面积为22 mm×11 mm。