This paper presents a comparison between finger and neck photoplethysmography (PPG) in order to assess the potential and limitations of this, non-conventionally used, body site for application in pulse oximetry. PPG signals were recorded at both sites from healthy subjects to inspect the differences in average waveforms, as well as in oxygen saturation (SpO2) and heart rate (HR) estimation. The results show significant differences in the average PPG pulse waveforms for different contour features such as diastolic or dicrotic notch amplitude, among others. The results show that the HR estimated from signals obtained with the neck sensor are strongly correlated to the output of the reference finger (R=0.862, MAE=1.27 BPM), whereas SpO2 measurements are not that accurately predicted (R=0.129, MAE=11.7%). Spectrograms under different breathing conditions revealed that the respiratory frequency is more predominant in neck PPG than in finger, which has a great potential for respiratory rate (RR) extraction. These are very promising results for the suitability of the neck as an alternative location for monitoring of respiratory diseases, and specifically for sleep apnea.
机译:本文介绍了手指和颈部光体积描记术(PPG)之间的比较,以评估这种非常规使用的身体部位在脉搏血氧仪中的潜力和局限性。在健康受试者的两个部位都记录了PPG信号,以检查平均波形以及血氧饱和度(SpO)的差异。 2 )和心率(HR)估算值。结果显示,在不同轮廓特征(例如舒张或重死切迹幅度等)的平均PPG脉冲波形中,存在显着差异。结果表明,根据颈部传感器获得的信号估算出的HR与参考手指的输出高度相关(R = 0.862,MAE = 1.27 BPM),而SpO 2 测量结果不能准确预测(R = 0.129,MAE = 11.7%)。在不同呼吸条件下的频谱图显示,颈部PPG中的呼吸频率比手指中的呼吸频率更为重要,这具有很大的呼吸频率(RR)提取潜力。这些对于颈部是否适合作为监测呼吸系统疾病,特别是睡眠呼吸暂停的替代部位的适用性,是非常有希望的结果。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号