This study evaluated the accuracy of ECG recordings captured by CardioChip - a single channel, low power, miniature ECG system-on-chip (SoC) designed for mobile applications, through comparison to a FDA cleared, widely-used standard 12-lead ECG recording device, Philips Pagewriter Trim III, within Mackay Memorial Hospital at Taiwan. There were 60 participants aged 39-87 evaluated. Two cardiovascular physicians found the ECG morphology to be the same between two devices, though CardioChip ECG showed more sensitivity to baseline noise. R-peak amplitudes measured by two devices showed statistical consistency. Offline analysis of signal correlation coefficients and coherence, both over 0.94 in average, indicated excellent correlation between two ECG recordings. The results suggested that CardioChip ECG is comparable to medical standard ECG.
机译:这项研究通过与FDA批准的,广泛使用的标准12导联ECG记录进行比较,评估了CardioChip捕获的ECG记录的准确性-一种用于移动应用的单通道,低功耗,微型ECG片上系统(SoC)设备是位于台湾麦凯纪念医院的飞利浦Pagewriter Trim III。有60位年龄在39-87岁之间的参与者进行了评估。两名心血管医师发现,两种设备之间的ECG形态相同,尽管CardioChip ECG对基线噪声显示出更高的敏感性。由两个设备测量的R峰幅度显示出统计一致性。信号相关系数和相干性的离线分析均均超过0.94,表明两个ECG记录之间具有极好的相关性。结果表明,CardioChip ECG可与医学标准ECG媲美。