首页> 外文会议>International Congress on Image and Signal Processing, BioMedical Engineering and Informatics >Automated Winston-Lutz Test for Efficient Quality Control in Stereotactic Radiosurgery
【24h】

Automated Winston-Lutz Test for Efficient Quality Control in Stereotactic Radiosurgery

机译:自动化温泉-LUTZ测试,以便在立体定向放射外科术中的高效质量控制

获取原文

摘要

Modern oncology and neurosurgery treatments often use stereotactic radiosurgery. Using linear accelerators and appropriate collimators, treatments of predefined high localized radiation doses can be concentrated in small volumes inside the patient. This can be used to treat tumors in otherwise difficult reachable zones of the patient, such as inside the head. In order to avoid unnecessary damage to surrounding healthy tissue, the treatment volume must be positioned as accurate as possible. The Winston-Lutz test is a standard test that is used in quality control procedures to verify the correct alignment of CT scan images with respect to the isocenter of the radiation equipment. The Winston-Lutz test uses a metal Tungsten ball positioned at the isocenter of the linear accelerator to calibrate the system under various positions of the gantry and various angles of the collimator in the gantry head and patient table. In current procedures, the calibration images that are captured during the test are manually interpreted, measured on workstation screens by the medical personnel. This paper presents new automatic methods to accurately determine the center of the metal ball, based on the image center of gravity as well as the automatic detection of the center of a square radiation field around the ball. The calibration results obtained from the automatic image processing based calculations are consistent and precise with a mean difference of 0.17 mm and a standard deviation of 0.15 mm.
机译:现代肿瘤和神经外科治疗通常使用立体定向放射外科手术。使用线性加速器和适当的准直器,可以在患者内部的小体积中浓缩预定义的高局部辐射剂量的处理。这可用于在患者的其他难以到达的区域中治疗肿瘤,例如头部。为了避免对周围健康组织的不必要损伤,必须尽可能准确地定位处理量。 Winston-Lutz测试是一个标准测试,用于质量控制程序,以验证CT扫描图像是否相对于辐射设备的Isocenter的正确对准。 Winston-Lutz测试使用位于线性加速器的Isocenter的金属钨球,以在龙门头和患者表中的各个位置处校准系统的各个位置和准直器的各种角度。在当前过程中,在测试期间捕获的校准图像被手动解释,在医务人员的工作站屏幕上测量。本文介绍了新的自动方法,可以基于手持的图像中心准确地确定金属球的中心。以及球周围的方形辐射场中心的自动检测。从自动图像处理的计算获得的校准结果是一致的,精确的平均差异为0.17 mm,标准偏差为0.15mm。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号