首页> 外文会议>International Conference on Biomedical Engineering >Multi-function electrical safety tester for medical equipment application
【24h】

Multi-function electrical safety tester for medical equipment application

机译:用于医疗设备的多功能电气安全测试仪

获取原文

摘要

The propose of this study was to design and construction of multi-function electrical safety tester for medical equipment application which used to measure the voltage, current from power supply, ground resistance and leakage current of the medical equipment. The multifunction electrical safety tester for medical equipment application was composed of four parts: 1) the measurement part consisted of voltage measurement circuit, current sensor, ground resistance measurement circuit and AAMI standard leakage current circuit, 2) the signal conditioning part consisted of voltage divider for voltage measurement, precision amplifier for leakage current measurement, non-inverting amplifier for ground resistance measurement and leakage current measurement, 3) the processing part consisted of Arduino Due microcontroller and 4) the display part consisted of liquid crystal display to show the amount of electricity that can be measured. The results of functional test of the research compared with the electrical safety analyzer (Fluke model ESA620) showed that the average percentage error of the voltage, electric current, ground resistance and leakage current were 0.47%, 2.37%, 0.67% and 1.17% respectively.
机译:这项研究的目的是设计和构造用于医疗设备的多功能电气安全测试仪,该测试仪用于测量医疗设备的电压,电源电流,接地电阻和泄漏电流。用于医疗设备的多功能电气安全测试仪由四个部分组成:1)测量部分由电压测量电路,电流传感器,接地电阻测量电路和AAMI标准泄漏电流电路组成; 2)信号调节部分由分压器组成用于电压测量,用于泄漏电流测量的精密放大器,用于接地电阻测量和泄漏电流测量的同相放大器,3)处理部分由Arduino Due微控制器组成,并且4)显示部分由液晶显示器组成,以显示可以测量的电量。与电气安全分析仪(Fluke ESA620型)相比,研究的功能测试结果表明,电压,电流,接地电阻和泄漏电流的平均百分比误差分别为0.47%,2.37%,0.67%和1.17% 。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号