Determination of blackbody cavity reflectance at 633nm by integrating sphere reflectometer

机译:积分球反射仪测定633nm黑体腔反射率

获取原文

摘要

The blackbody cavity emissivity is a primary parameter of blackbody source. The measurement of blackbody cavity emissivity could verify the design value of cavity emissivity. A determination of graphite blackbody cavity emissivity by reflectance measurement is presented. The measuring equipment include a He-Ne laser, laser power stabilizer, integrating sphere, Si photoelectric detector, gray reflectance standard sample and Keithley 6485 picoammeter. The preliminary measurement value of graphite blackbody cavity emissivity is 0.99983351.
机译:黑体腔发射率是黑体源的主要参数。黑体腔发射率的测量可以验证腔发射率的设计值。提出了通过反射率测量确定石墨黑体腔发射率的方法。测量设备包括氦氖激光器,激光功率稳定器,积分球,硅光电探测器,灰度反射标准样品和吉时利6485皮安计。石墨黑体腔发射率的初步测量值为0.99983351。