Mechanical Properties of Polypropylene-ePTFE Surgical Mesh Explanted After in vivo Function

机译:体内功能植入聚丙烯-ePTFE手术网的力学性能

获取原文

摘要

These preliminary results indicate that COMPOSIX™ EX meshes are stiffer after relatively short duration of exposure to the physiological environment. Previous studies have noted similar stiffening after in vivo function. Costello found between a 4 and 30 fold decrease in the compliance of explanted meshes. The underlying cause of such increased stiffness and its relationship with clinical symptoms remains to be determined. Testing of the remaining samples in the registry of explanted surgical mesh, including both COMPOSIX™ EX and other brands, is ongoing.
机译:这些初步结果表明,COMPOSIX™EX筛网在暴露于生理环境中的时间相对较短后会变硬。先前的研究已经注意到体内功能后类似的硬化。 Costello发现外植网的顺应性降低了4到30倍。这种硬度增加的根本原因及其与临床症状的关系尚待确定。正在进行对移植手术网注册表中剩余样品的测试,包括COMPOSIX™EX和其他品牌。