Electromagnetic techniques to minimize the risk of hazardous local heating around medical implant electrodes during MRI scanning

机译:电磁技术可最大程度减少MRI扫描过程中医疗植入电极周围危险的局部加热风险

获取原文

摘要

Magnetic Resonance Imaging (MRI) scans are contraindicated for many patients with medical implants. We establish the circumstances that cause, and the resistances required to ameliorate and to eliminate dangerous levels of MRI-induced heating that occur at the exposed, distal end of an electrical lead implanted in tissue. Simulated predictions are compared with measurements made at 128 MHz in a 3-Tesla MRI machine. A low resistance at kilohertz frequencies is sought by implant makers, in contrast with the high resistance demanded for safety. The practicality of presently-developed strategies to prevent tissue damage is brought into question. We examine the extent to which skin-depth and transmission-line properties can be manipulated to improve safety.
机译:磁共振成像(MRI)扫描禁止用于许多有医疗植入物的患者。我们确定了造成这种情况的原因,以及改善和消除在组织中植入的电引线的裸露远端发生的MRI感应加热的危险水平所需的电阻。将模拟预测与在3-Tesla MRI机器中以128 MHz进行的测量进行比较。与安全性要求的高电阻相反,植入物制造商寻求在千赫兹频率下的低电阻。当前开发的防止组织损伤的策略的实用性受到质疑。我们研究了可操纵集肤深度和传输线属性以提高安全性的程度。